Philips RESPIRONICS SimplyGo Mini Mode D'emploi page 50

Masquer les pouces Voir aussi pour RESPIRONICS SimplyGo Mini:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Electrical Fast
±2 kV for Power Supply
Transient/Burst
Lines
IEC 61000-4-4
±1 kV for Input/Output
Lines
Electrical Transient
As specified in
Conduction along
ISO 7637-2
supply lines
ISO 7637-2
Surge
±1 kV Line to Line
IEC 61000-4-5
±2 kV Line to Ground
Voltage Dips, Short
<5% U
Interruptions and
U
Voltage Variations on
degree increments
Power Supply Input
Lines
70 % U
IEC 61000-4-11
for 0�5 seconds
<5% U
for 5 Seconds
Power Frequency
30 A/m
(50/60 Hz) Magnetic
Field
IEC 61000-4-8
Note: U
is the A�C� mains voltage prior to application of the test level�
T
Guidance and Manufacturer's Declaration – Electromagnetic Immunity
SimplyGo Mini is intended for use in the electromagnetic environment specified below� The user of SimplyGo Mini should assure
that it is used in such an environment�
IEC 60601
Immunity Test
Test Level
Conducted RF
3 Vrms
IEC 61000-4-6
150 kHz to
80 MHz
Radiated RF
6 Vrms
IEC 61000-4-3
Amateur
Radio &
ISM Bands
between 150
kHz and 80
MHz
Radiated RF
10 V/m
IEC 61000-4-3
80 MHz to
2�7 GHz
AMSR
±2 kV for Power Supply
Lines
±1 kV for Input/Output
Lines
As specified in
ISO 7637-2
±1 kV Line to Line
±2 kV Line to Ground
(>95% Dip in
<5% U
T
) for 0�5 Cycle at 45
for 0�5 Cycle at 45 degree
T
increments
(30% Dip in U
)
70 % U
T
T
0�5 seconds
(>95% Dip in U
)
<5 % U
T
T
for 5 Seconds
30 A/m
Compliance
Level
Electromagnetic Environment - Guidance
Portable and mobile RF communications equipment should be used no closer
to any part of the Device, including cables, than the recommended 30 cm
separation distance�
3 Vrms
150 kHz to 80
MHz
6 Vrms
Amateur Radio
& ISM Bands
between 150
kHz and 80 MHz
10 V/m
Interference may occur in the vicinity of equipment marked
with the following symbol:
oxygenconcentratorstore.com
SimplyGo Mini User Manual
Mains power quality should be that of a typical
home or hospital environment�
The Device is suitable for use in vehicles and
boats�
Mains power quality should be that of a typical
home or hospital environment�
(>95% Dip in U
)
Mains power quality should be that of a typical
T
T
home or hospital environment� If the user of the
Device required continued operation during
power mains interruptions, it is recommended
(30% Dip in U
) for
that the Device be powered from an
T
T
uninterruptible power supply or battery�
(>95% Dip in U
)
T
T
Power frequency magnetic fields should be at
levels characteristic of a typical location in a
typical home or hospital environment�
1-877-774-9271
47

Publicité

Table des Matières
loading

Table des Matières