Note: To Equipment Provider: The following oxygen
administration accessories are recommended for use
with the Companion 5:
• Nasal Cannula: CAIRE Part Number CU002-1
• Humidifier Adaptor Tubing: CAIRE Part number
20843882
• Humidifier Bottle: CAIRE Part Number HU003-1
• Firebreak: CAIRE Part Number 20629671
A firebreak is required for use with any cannula.
• CAIRE offers a firebreak intended to be used
in conjunction with the oxygen concentrator. The
firebreak is a thermal fuse to stop the flow of gas in
the event that the downstream cannula or oxygen
tubing is ignited and burns to the firebreak. It is
placed in-line with the nasal cannula or oxygen tubing
between the patient and the oxygen outlet of the
Companion 5. For proper use of the firebreak, always
refer to the manufacturer's instructions (included with
each firebreak kit).
• Additional recommended accessories information is
available online at www.caireinc.com.
PN 21420526-C4 B | User Manual 13 - ENG