Télécharger Imprimer la page

Hologic Aquilex Instructions D'utilisation page 51

Masquer les pouces Voir aussi pour Aquilex:

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 125
Technical Data
en
Accuracy
Dimensions
Weight [kg], [lbs]
Interfaces:
Essential Performance
50 / 378
Deficit [ml]
-995/+9995 ml
Pressure [mmHg]
± 10 mmHg
< 1 l: ± 60 ml,
Deficit [ml]
> 1 l: ± 6 % deficit accuracy related to provided inflow volume, but maximum
300 ml
Suction pressure
± 20 %
Total inflow volume [ml]
± 10 %
Width x Height x Depth
300 mm x 140 mm x 300 mm / 11.8 in x 5.5 in x 11.8 in (AQL-100PBS),
[mm], [in]
670 mm x 1390 mm x 670 mm / 26.4 in x 54.8 in x 26.4 in (AQL-100CBS)
5.8 kg, 12.8 lbs (AQL-100PBS)
14.2 kg, 31.3 lbs (AQL-100CBS)
Irrigation Pump Unit:
1 x scale connection (flanged socket/round connection socket with 5 pins)
1 x service port (RS232 socket DSUB9/D-SUB9)
Fluid Monitoring Unit/Container Scale:
ON/OFF signal for compon-
1 x data connection (flanged female / round socket with 5 pins) for connection
ents
to bag scale.
Fluid Monitoring Unit/Bag Scale:
2 x data ports (flanged socket/round connection socket with 5 pins): Con-
nector to connect to pump, socket to connect with container scale
Mains power socket
IEC 60320-1 C14
Working frequency: 13.5609 MHz,
RFID Transponder Techno-
logy
Transmitting power: -7.51 dBμA/m at 10 m
Pressure build-up in the body cavity, control and measurement, limit: max.
150  mmHg (Normal Condition), 200  mmHg for max. 5  seconds (Single Fault
Condition).
Suction Pressure: up to -67  kPa (Normal Condition), 0  kPa by no function or
-84 kPa by limit due to design constraints (Single Fault Condition)
Deficit measurement:
< 1 l: ± 60 ml,
> 1 l: ± 6 % deficit accuracy related to provided inflow volume, but maximum
300 ml
Single Fault Condition 1: no function of deficit measurement
Single Fault Condition 2: < 1 l: greater deviation than ± 60 ml, > 1 l: greater de-
viation than ± 6 % deficit accuracy related to provided inflow volume or more
than 300 ml
1
all the technical requirements for an application part.
A tube set is not an application part in terms of the standard. However, it meets

Publicité