Function Test

WARNING!
Function test
The function test must be performed prior to each device use.
6.1

General Device and Setup Check

1.
Perform a visual check of the devices. Do not use the system in case of obvious
damage.
2.
Check the rollers of the roller wheel to make sure they move easily and smoothly.
3.
Turn on the device, check whether power switch and indicators and displays
light up accordingly.
4.
The device self-test must run successfully; there are no displayed error mes-
sages (see Section Turning On the Aquilex® System [} 28]).
5.
The bags with the irrigation fluid must hang freely and may not touch the scale.
6.
Check that all tube connections (vacuum/inflow/outflow) were made correctly
and are intact.
7.
Check to make sure all tube connections are free of mechanical stresses and
are routed without snagging. The tube connections may not touch the scale.
Non-observance can lead to a distortion of the deficit calculation.
8.
The automatic lumen calibration has been carried out successfully; there are no
displayed error messages (see Section Using the Pump during Surgery [} 35]).
9.
Check for leaking irrigation fluid in the area of the pressure chamber.
6.2

Scale Test

You need the following for this test: 1000 g weight (included with each pump).
1.
Switch system on.
2.
Once the display depicts Insert Tube Set, press Pause/Resume and Zero at the
same time.
3.
The pump display depicts the message Scale Test.
The first option is the test of the container scale.
4.
Place the 1000 g weight on the container scale.
5.
The displayed value should be 1000 g. The tolerance range is ±20 g.
6.
If the displayed value is outside the tolerance range, the scale must be calib-
rated by a service technician.
7.
Remove weight from container scale.
Press the º button (see Fig. Front of Irrigation Pump Unit [} 22], (11) Increase
8.
the deficit limit) to start the test of the bag scale.
9.
The test of the bag scale must be carried out on both sides of the bag scale (us-
ing left and right bag hook).
10. Place the 1000 g weight on the right bag hook of the bag scale.
11. The display of the fluid deficit limit depicts the weight. The displayed value
should be 1000 g. The tolerance range is ±20 g.
12. If the displayed value is outside the tolerance range, the scale must be calib-
rated by a service technician.
13. Repeat the steps 10. to 12. with using the left bag hook of the bag scale.
14. Remove weight from bag scale.
(Switch to the container scale test by pressing the » button (see Fig. Front of Ir-
rigation Pump Unit [} 22], (10) Decrease deficit limit).)
15. Press the Pause/Resume button to end the test.
Enter the results into the test log in section Test log [} 56]. The test is successfully
completed when the results are within the permissible tolerance limit.