WARNING!
Obvious defects
Never use the system if it has suspected or confirmed defects, especially if these in-
volve the power plugs or the mains power supply connection cables. In this case
have the device repaired by authorized service personnel.
5.5

Hanging the Fluid Bags

WARNING!
Distension media
When performing monopolar hysteroscopic electrosurgery, the distension medium
must be electrically non-conductive. Examples include glycine, sorbitol and man-
nitol. Isotonic saline irrigation fluids may only be used when performing bipolar
electrosurgical resective procedures.
Hang one or two fluid bags filled with distension fluid appropriate for procedure.
WARNING!
Irrigation fluid bags
The system is only intended for use with flexible fluid bags. Do not use glass con-
tainers as they might break. With rigid containers, fluid cannot flow quickly
enough due to the vacuum being generated inside of the containers. Risk of implo-
sion with rigid containers.
Fig. 5–8  Fluid bag suspension
(1)
(2) Fluid bag on bag hook
(1)
(3) Bag deflector
(2)
(3)
System Operation
Bag scale
29 / 378