Limit Value, Hysteresis, Contact Type; Icons In The Measurement Display - Burkert 8285 El. COND Mode D'emploi

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Limit value

Limit value, hysteresis, contact type

Parameter setting/Module BASE/Relay contacts/Usage
Menu Display
Contact K1 (Administrator)
Usage
Variable
Limit value
Hysteresis
Effective direction
Contact type
Return
Limit value
Effective direction min
Hysteresis +
Limit value
1
Contact
0

Icons in the measurement display:

Measured value exceeds limit:
Hysteresis
Tolerance band around the limit value, within which the contact is not actua-
ted. Serves to obtain appropriate switching behavior at the output and
suppress slight fluctuations of the measured variable (Fig.)
Contact type
Specifies whether the active contact is closed (N/O) or open (N/C).
7.00 pH
19.2°C
NAMUR maintenance
NAMUR function check
Limit value
Rinse contact
Parameter set 2 active
USP output
Signal
Usage as limit value
Relay output: Limit value
Stromausgang parametrieren
• Parametrierung aufrufen
Call up parameter setting
• Paßzahl eingeben
Enter passcode
• Modul BASE wählen
Select "Module BASE"
• "Augangsstrom I1" auswählen
Select "Contact ..."
"Usage: Limit" (Fig.)
Limit value
Effective direction max
Limit value
Hysteresis -
1
Contact
0
Measured value falls below limit:
Signal
65

Publicité

Table des Matières
loading

Table des Matières