Menu Display
Menu selection
Return to meas
Calibration
Module 8285 El. COND
Module 8285 pH/ORP
Module 8285 El. COND
Return
Module 8285 El. COND
Automatic with standard cal solution
Manual entry of cal solution
Product calibration
Data entry - premeasured sensor
Return
28
0.020 mS/cm
25.6 °C
Select:
[enter]
Lingua
0.020 mS/cm
25.0°C
Info
0.020 mS/cm
25.0 °C
Select calibration method
Call up calibration
Press menu key to select menu.
Select calibration using arrow keys,
confirm with enter, passcode 1147
(The passcode can be edited by the
administrator.) After passcode entry,
the analyzer is in function check
mode: Current outputs and relay
contacts behave as configured until
the Calibration menu is exited.
Calibration:
Select "Module COND"
Select calibration method:
• Automatic with standard cal solution
• Manual entry of cal solution
• Product calibration
• Data entry - premeasured sensor
When you call up calibration, the
analyzer automatically proposes the
previous calibration method.
If you do not want to calibrate,
"Return" with the left softkey.
8285