Concentration Chart (Add. Function) - Burkert 8285 El. COND Mode D'emploi

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Concentration chart (add. function)

Menu selection: Parameter setting/System control/Concentration chart
Specifying a concentration solution for conductivity measurement
Concentration chart (additional function SW 8285-009)
To specify the customer-specific solution, 5 concentration values A-E are
entered in a matrix together with 5 temperature values 1-5. To do so, first
enter the 5 temperature values, then enter the respective conductivity values
for each concentration A-E.
These solutions will then be available in addition to the permanently stored
standard solutions (select "Chart").
Menu Display
System control (Administrator)
Calculation Blocks
Tag number
Release of options
Logbook
Concentration table
Factory setting
Return
Concentration chart (Administrator)
Temperature 1
Temperature 2
Temperature 3
Temperature 4
Temperature 5
Abort
Concentration chart (Administrator)
Concentration A:
1. Cond at +000.0 °C
2. Cond at +005.0 °C
3. Cond at +010.0 °C
4. Cond at +015.0 °C
5. Cond at +020.0 °C
Abort
The concentration chart is selected as follows:
Parameter setting / Module COND / Concentration = ON / Medium = Chart.
0.020 mS/cm
25.6 °C
0.020 mS/cm
25.6 °C
+000.0 °C
+005.0 °C
+010.0 °C
+015.0 °C
+020.0 °C
OK
0.020 mS/cm
25.6 °C
05.00 % by wt
0.000 µS/cm
0.000 µS/cm
0.000 µS/cm
0.000 µS/cm
0.000 µS/cm
OK
Enter concentration chart
Enter values
• Call up parameter setting
• System control
• Select "Concentration chart"
Enter 5 temperature values
(right/left arrow keys to
select position, up/down arrow to
edit number, confirm with enter.)
Enter values for concentrations A-E
for the respective temperatures. The
table values must be continuous.
Maxima/minima are not permitted.
Incorrect entries are marked with .
47