Temperature compensation
Calibration / Adjustment
Temperature compensation
Temperature compensation during calibration
The conductivity value of the calibration solution is temperature-depen-
dent. For calibration, the calibration solution temperature must therefore be
known in order to choose the actual value from the conductivity table.
During parameter setting you define whether cal temperature is measured
automatically or must be entered manually.
Automatic temperature compensation
Calimatic
Dip sensor in cal solution
then "Start" calibration
Cal solution
TC automatically considered
Measured cal temp
Sensor replacement
Return
Otherwise, you must select manual entry of calibration temperature. When
"Cal temp automatic" is set, "Measured cal temp" appears in the menu.

Manual temperature compensation

Calimatic
Dip sensor in cal solution
then "Start" calibration
Cal solution
TC automatically considered
Enter cal temp
Sensor replacement
Return
26
For automatic cal temp detection, the analyzer
0249 mS/cm
25.6 °C
measures the temperature of the calibration
solution with a temperature probe (Pt 100 /
Pt 1000 / NTC 30 kΩ). If you work with auto-
NaCl saturated
matic temperature compensation during cali-
+025.6 °C
bration, a temperature probe connected to
the temperature input of the 8285 El. COND
Proceed
module must be in the calibration solution.
The temperature of the calibration solution
0249 mS/cm
25.6 °C
must be entered manually in the Calibration
menu. When "Cal temp automatic" is set,
"Measured cal temp" appears in the menu.
NaCl saturated
When "Cal temp manual" is set,
+025.6 °C
"Enter cal temp" appears in the menu.
Proceed
8285