Parameter Setting; Default Settings And Selection Range; Tc Process Medium - Burkert 8285 El. COND Mode D'emploi

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Parameter setting

Default settings and selection range

Note: Function check active
Parameter
Input filter
• Pulse suppression
Sensor data
• Sensor type
• Nominal cell constant
• Temperature detection
Measuring temp
Cal temp
• Sensocheck
Cal preset values
• Calibration solution
• Product calibration

TC process medium

• TC correction
• Reference temp
• Impurity
42
Default
Selection / Range
Off, On
Off
2-El sensor, 4-El sensor
2-EL sensor
X.XXXX (entry) depending on selected sensor
1.0000 cm*
Pt100, Pt1000, NTC30kOhm, Ni100 (sensor selection)
(Pt1000)
Auto
Auto, manual: Default +25.0 °C (entry)
Auto, manual: Default +25.0 °C (entry)
Auto
Off, failure, maintenance request
Off
KCl 0.01 mol/l
without TC
Off
KCl 0.01 mol/l (1.413 mS/cm)
KCl 0.1 mol/l (12.88 mS/cm)
KCl 1 mol/l (111.80 mS/cm)
NaCl 0.01 mol/l (1.183 mS/cm)
NaCl 0.1 mol/l (10.683 mS/cm)
NaCl saturated (251.3 mS/cm)
without TC, with TC
Off, linear, EN 27888, ultrapure water
(Linear: Enter reference temp +025.0 °C)
(Ultrapure water: NaOH, NaCl, HCl, NH 3 )
(Adjustment range depending on parameter)
8285

Publicité

Table des Matières
loading

Table des Matières