6. Care and Maintenance

6.3 Changing the Air Filter

When the message "Replace Air Filter" appears, the HEPA filter is in need of changing.
To continue using the Hydrim, press OK. Call for service as soon as possible. The air filter
should be changed before 750 cycles.
salt container
6.2 Refilling the Water Softening

Salt Reservoir

When the error message "salt level low" appears,
the salt reservoir is in need of refilling. Ensure that
your authorised technician has set the water softener
according to the hardness level in your area. If the
level is set at 0, no message will appear and salt is
not required.
To fill the salt reservoir, follow these steps:
1. Unscrew the salt container lid.
2. Fill the salt container to the top
(maximum 1 kg / 2.2 lbs.)
3. Screw the salt container lid on tightly.
4. Run a complete cycle, with no instruments.