6. Care and Maintenance

6.3 Cleaning the Hydrim L110w

To clean outer surfaces and the decal covering the touchscreen, use a soft cloth
moistened with soap and water or a surface disinfectant. Do not use harsh cleaning
chemicals.
If preparing the unit for transport or storage, waste water should be removed
from the sump.
Waste water in the unit may contain biological contaminants. Use a mechanical means
to siphon the contents. Wear disposable rubber gloves. Dispose of absorbent material
according to biological waste disposal regulations.
salt container
6.2 refilling the Water Softening

Salt reservoir

When the error message "salt level low" appears,
the salt reservoir is in need of refilling. Ensure that
your authorized technician has set the water softener
according to the hardness level in your area.
To fill the salt reservoir, follow these steps:
1. Unscrew the salt container lid.
2. Fill the salt container to the top
(maximum 1 kg / 2.2 lbs.)
3. Screw the salt container lid on tightly.
4. Run a complete cycle, with no instruments.