7. Troubleshooting
Error Messages
Error
CF1 Water Heating
failure
CF2 Chamber
Filling Failure
CF3 Chamber
temperature
reading failure
CF4 draining
failure
CF9 Software or
PCB Failure
CF11 No detergent
CF14 Fault Flow Switch
CF15 Chamber
overflow
Cause
The water does not reach the
required temperature in the
specified time
The water does not fill the
chamber within the specified time.
Faulty sensor.
Water cannot drain
from the unit.
Cycle length is more than
2hr 30 min.
Flow switch did not detect detergent
during wash phase.
Flow switch is stuck closed.
Overflow switch did not turn off after
30 sec of running the drain pump.
22
Solution
Call for service.
Check that the water hoses
are not kinked. Open the
water shut-off valves.
Call for service.
Check that the drain hose is
not kinked. Check that the
filter in the bottom of the
chamber is not obstructed.
Call for service.
Replace detergent box
if empty.
If the problem persists,
call for service.
Call for service.
Check if drain is blocked;
Call for service.