4. installation instructions

Connecting the water inlet hoses:

Connect the hot and cold water inlet hoses to the hot and
cold water supply.
drain requirements:
Connect the drain hose to the drain outlet. The drain
hose can be attached to existing drain lines using a
3.5 cm / 1.5" or larger standpipe / P-trap combination. If
the hose is connected directly to the drain line, fittings
and adapters should not reduce water flow. The drain
hose should be attached to the main drain at a point no
more than 1 metre /3 ft. above the base of the Hydrim.
A floor drain is acceptable (check local codes).

Electrical requirements:

Connect the electrical cord to the power supply. The
Hydrim must have a dedicated circuit equipped with a
15A fuse. The outlet must be grounded.

Additional information:

• The Hydrim unit is heavy. Use a forklift or dolly and
exercise caution when moving it.
• The Hydrim should be serviced on site.
• The Hydrim is equipped with an air gap / anti-suction
device to prevent backflow of dirty water into the
water supply. No other air gap device is necessary.
• If you need to extend the water inlet and drain hoses,
ensure that you use commercial grade plumbing hose.
The maximum length of the drain hose is 3.3 m / 13 ft.
installing Cleaning Solution:
Install HIP Cleaning Solution as follows:
installation Test:
T urn on the shut-off valves. Run a cycle, checking for leaks in the plumbing connections.
Testing water hardness and
setting the water softener (salt):
Hydrim is equipped with a built-in water softening
system which needs to be adjusted according to
the local water hardness. The Hydrim water test kit
includes three water hardness test strips in bags. Take
a water sample from the location where the machine
will be installed. Open one of the bags and remove
the test strip. Dip the strip in the water. Compare the
color of the strip with the chart on the back of the bag.
Determine the water hardness according to the chart
on the water test kit envelope. Power the unit on.
Touch the "i" in the lower right hand corner of the
sceen. Select
"Setup" and then
"Set Regeneration".
Using the up and
down arrows, set
the water softener
regeneration level
according to this
table:
consider using an addition-
*
al water treatment
Unscrew the salt
container lid and pour
1 litre / 1 quart of water
into the water softener.
Add 1 kg / 2.2 lbs of
water softening salt in
the same manner.
Screw the salt container
cap on tightly.
5
Hardness – ppm
Hydrim setting
0 - 180
190 - 210
220 - 250
260 - 300
310 - 350
360 - 400
410 - 540
550 - 890*
> 890
treatment required
salt container
0
1
2
3
4
5
6
7
water