Refilling The Water Softener Salt Reservoir; Changing The Air Filter; Cleaning The Hydrim M2 G4 - SciCan HYDRIM M2 G4 Mode D'emploi

Masquer les pouces Voir aussi pour HYDRIM M2 G4:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

7. Care and Maintenance (cont.)

7.2.2 Refilling the Water Softener salt reservoir

When the error message "salt level low" or the
water softener/detergent icon indicated by a red X
appears, the salt reservoir is in need of filling.
To fill the salt reservoir proceed as follows:
1. Unscrew the salt container lid.
2. Fill the salt container to the top
(maximum 1 kg / 2.2 lbs.)
3. Screw the salt container lid on
tightly.
4. Run a complete cycle, with no
instruments.
Important notes:
If the salt container is being filled for the first time, it MUST be filled with
at least 1 litre of water before adding the salt.
Only tablet salt should be used in the softening unit. Fine grained salt
such as table salt should NOT be used as it could clog the unit.
The water softener salt is also available from SciCan. Please contact your
local dealer.

7.3 Changing the Air Filter

When the message "Replace Air Filter" appears, the HEPA filter is in need of changing.
To continue using the HYDRIM, press OK. Call for service as soon as possible. The air
filter should be changed at 600 cycles.

7.4 Cleaning the HYDRIM M2 G4

To clean outer surfaces and the decal covering on the touchscreen, use a soft cloth
moistened with soap and water or a surface disinfectant. Do not use harsh cleaning
chemicals.
If preparing the unit for transport or storage, waste water should be removed from
the sump.
Waste water in the unit may contain biological contaminants. Use a mechanical means
to siphon the contents. Wear disposable rubber gloves. Dispose of absorbent material
according to biological waste disposal regulations.
salt container
35

Publicité

Table des Matières
loading

Table des Matières