5. Instructions for Use (cont.)
5.4 Recommended load configurations.
The intended purpose of the M2 G4 washer-disinfector is a clinical general instrument
washer-disinfector, predominantly aimed at the general dental and medical markets.
There will be many combinations of instruments that can be processed in the M2 G4 unit
depending on:
• Size of practice
• Types of procedures undertaken at any one time
• Specialization of practice
• Instrument management protocols
It is difficult to illustrate all the possible combinations av ailable, so below we have given
examples of the most common typical basket loadings, the recommended loading
of general instruments, and the worst case loading, for which the unit has been type
tested and validated.
Handle all instruments with care to prevent personal injuries through
punctures for example. Follow local health and safety regulations for
preventing accidents, bodily harm and personal injuries.

5.4.1 Baskets and Accessories

Observe the instrument manufacturer's recommendations for cleaning and
care. When loading the unit, arrange loose instruments so that they do not
touch one another. Instruments with concave surfaces or cavities should be
placed so that water can drain freely.
Sample loading:
Example 1 – solid straight & hinged instruments
Example 2 – solid straight instruments
For example:
scalpels, scissors,
curettes, excavators,
pluggers, etc.
For example:
pliers, knives,
explorers, scalers,
etc.
13