5. Instructions for Use (cont.)
5.4.2 Type tested loads.
Shown below are two load configurations used for the type testing of the M2 G4 unit.
The M2 G4 has been validated for the following reference loads. These two loads
represent commonly used loads in a dental office, although many other load
combinations are possible. In cases where loads differ from the reference loads
an initial on-site validation by the installation technician may be required and is
recommended. If one of these factory validated loads is used by the dental office and
this can be documented by pictures and/or appropriate lists, an initial onsite validation
may not be necessary.
Type test load 1
Hinged instruments, solid instruments and
open, hollow instruments.
• 2 x 01-110411S - vertical instrument rack
• 2 x 01-108232S - long basket
Type test load 2
Instrument sets in cassettes - full load
• Top trolley - 1 x 01-109963S - cassette rack,
full size + 5 x 8x11x1 instrument cassettes
• Bottom trolley - 1 x 01-109963S - cassette rack,
full size + 5 x 8x11x1 instrument cassettes