SciCan HYDRIM M2 G4 Mode D'emploi page 34

Masquer les pouces Voir aussi pour HYDRIM M2 G4:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

7. Care and Maintenance
5.
If appropriate, remove any dried chemical by immersing the connector in a cup of water.
6.
Place the new container of solution on the solution door (see picture 3 below).
7.
Remove the tear-off portion of the outer carton. (See picture 4 below).
8.
Pull out the internal chemical bag nozzle and secure it in the cutout (see picture 5a).
Remove the cap.
9.
Connect the cleaning solution connector to the new solution container by screwing
clockwise until it stops rotating (see picture 5b). Caution – avoid cross threading the
connector as damage may occur and the joint may leak.
10.
Insert the new solution container by rotating forwards into the correct location in the
solution door (see picture 6).
11.
Close the cleaning solution door.
12.
Press OK on the touch screen. The machine display will indicate "Priming. Please wait".
13.
When the cycle selection menu appears, priming is complete and the machine is ready
for use.
14.
Proceed with the next cycle.
Picture 1
To prime the cleaning solution dosing pump, press the water softener/detergent icon on the
main screen. In the water softener/detergent screen, press the red X next to "Detergent".
The unit will prime the dosing system and a green check mark will appear in place of the
red X when it is ready for use.
NOTE: The system can also be primed by simply starting a cycle and selecting "Detergent
Replaced", when prompted.
Picture 2
Picture 3
Picture 4
Picture 5
34
Picture 6

Publicité

Table des Matières
loading

Table des Matières