5. Instructions for Use

5.1 Instrument Reprocessing Recommendations

The Hydrim M2 is designed to clean and thermally disinfect (depending on the cycle
chosen) general dental and medical instruments.
The disinfection level (P2 minimum A
reduce the number of viable microorganisms on a product to a level appropriate for its
intended further handling or use depending on requirements specified by regional or
national authorities. Regulatory authorities may specify more stringent requirements
within the territories for which they are responsible.
SciCan recommends terminal sterilisation after processing in the Hydrim M2.
When used correctly, the Hydrim M2 is effective for the removal of debris on
instruments that may result in the insulation of microorganisms during terminal
sterilisation. Prior to placing instruments in the Hydrim, consult the instrument
manufacturer's reprocessing instructions.
Examples of medical and dental instruments suitable for cleaning and disinfecting in
the Hydrim M2 include:
• Stainless steel scalers, spatulas and hemostats.
• Stainless steel retractors, needle holders, skin hooks, tissue forceps, scissors,
curettes, and other stainless steel instruments free of deep cavities.
For best results, instruments processed in the Hydrim M2 should
have the following properties:
• Heat resistance to a temperature of up to 96°C
• Corrosion resistance in the presence of heat and alkalinity.
Microkeratomes, phaco tubing, fiberoptics and electrical equipment should not be
processed in the Hydrim.
HIP
TM
(Hydrim Cleaning Solution with Instrument Protection) has been carefully
formulated to provide outstanding material compatibility with a wide range of metals
and coatings. However, some manufacturers' instruments may not be compatible.
When processing small or light objects in the Hydrim use the basket with lid
(SciCan part number 01-109966S).
= 600, P3 minimum A
0
37
= 3000) is intended to
0