5.10 Opening the Door in Case of Power Failure

If there is a power failure, ensure the machine is isolated from the main power supply.
Then remove the kickplate of the Hydrim. Remove the screw with a screwdriver and lift
the kickplate up and toward you. Locate the ring and pull it to open the door. Exercise
caution! There may be fluid remaining in the unit and the instruments may be hot.
Instruments that have not completed the cycle should not be used and should be
reprocessed. When replacing the kickplate, ensure that the grounding wire is attached.
Ring
Grounding wire
Kickplate
Screw