Opening The Door In Case Of Power Failure; Printer / Usb Setup - SciCan HYDRIM C51w Mode D'emploi

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

5. Instructions for Use

5.11 Printer / USB Setup

The HYDRIM C51w has an RS-232 port at the back, and can be used with an external
printer or the SciCan Data Logger. The printers in the chart below have been tested
with the HYDRIM. To add or change a printer or SciCan Data Logger, follow these steps:
Turn off the HYDRIM and the printer or Data Logger before connecting these devices
to the unit.
1.
With the printer or Data Logger connected, turn on the HYDRIM and press the i to
move to the Menu screen.
2.
In the Setup Menu, select Printer Selection.
3.
Select Serial Printer if connecting
a printer, or USB Flash/MSD if
connecting the SciCan Data Logger.
Press the back arrow to return to
the Setup Menu.
4.
In the Setup Menu, select Baud rate.
5.
Select the rate required (refer to
chart below for recommended Baud
rates). Use the back arrow to return
to the Start screen.
6.
Ensure that the date and time
have been set (see Setup Menu –
Date / Time).
7.
Now the HYDRIM will write its cycle
information to the device chosen.
Page 19

5.10 Opening the Door in Case of Power Failure

If there is a power failure, open the detergent door and remove
the chemical bottle. Keep the chemical bottle upright and
ensure that the chemical does not spill. Locate the ring in
the top of the detergent compartment and pull it to open the
door. Exercise caution! There may be fluid remaining in the
unit and the instruments may be hot. Instruments that have
not completed the cycle should not be used and should be
reprocessed.
Printer Model
Epson
TM-U220D
Citizen
IDP-3110-40 RF 115
Star Micro
SP212FD42-120
Star Micro
SP216FD42-120
Star Micro
SP512MD42-R
SciCan Data Logger
For Mass Storage Device
Serial Port
Baud Rate
9600
9600
9600
9600
9600
Serial Port
Baud Rate
9600

Publicité

Chapitres

Table des Matières
loading

Table des Matières