SciCan HYDRIM C51w Mode D'emploi page 26

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

7. Troubleshooting
Error Messages
Error
CF1 Water Heating
failure
CF2 Chamber
Filling Failure
CF3 Chamber
temper ature
reading failure
CF4 Water
evacuation
failure
CF9 Software or
PCB Failure
CF15 Chamber
overflow
Page 24
Cause
The water does not reach the
required temperature in the
specified time
The water does not fill the
chamber within the specified time.
Faulty sensor.
Water cannot drain
from the unit.
Cycle length is more than
2hr 30 min.
Overflow switch did not turn off
after 30 sec of running the drain
pump.
Solution
Call for service.
Check that the water hoses
are not kinked. Open the
water shut-off valves.
Call for service.
Check that the drain hose is
not kinked. Check that the
filters in the bottom of the
chamber are not obstructed.
Call for service.
Check if drain is blocked.
Call for service.

Publicité

Chapitres

Table des Matières
loading

Table des Matières