5.1 Instrument Reprocessing Recommendations

The Hydrim C51wd is designed to clean and thermally disinfect (depending on the cycle
chosen) general dental and medical instruments.
The disinfection level (A
on a product to a level appropriate for its intended further handling or use depending on
requirements specified by regional or national authorities. Regulatory authorities may
specify more stringent requirements within the territories for which they are responsible.
SciCan recommends terminal sterilisation after processing in the Hydrim C51wd.
When used correctly, the Hydrim C51wd is effective for the removal of debris on
instruments that may result in the insulation of microorganisms during terminal sterilisation.
Prior to placing instruments in the Hydrim, consult the instrument manufacturer's
reprocessing instructions.
Examples of medical and dental instruments suitable for cleaning and disinfecting in the
Hydrim C51wd include:
• Stainless steel scalers, spatulas and hemostats.
• Stainless steel retractors, needle holders, skin hooks, tissue forceps, scissors,
curettes, and other stainless steel instruments free of deep cavities.
For best results, instruments processed in the Hydrim C51wd should have the following
properties:
• Heat resistance to a temperature of up to 84°C
• Corrosion resistance in the presence of heat and alkalinity.
Dental handpieces, microkeratomes, phaco tubing, fiberoptics and electrical equipment
should not be processed in the Hydrim.
HIP
TM
(Hydrim Cleaning Solution with Instrument Protection) has been carefully formulated
to provide outstanding material compatibility with a wide range of metals and coatings.
However, some manufacturers' instruments may not be compatible.
All hinged instruments should be placed in open baskets in P2 Heavy Duty cycle or P3
Heavy-Duty with Disinfection cycle.
When processing small or light objects in the Hydrim use the basket with lid (SciCan part
number 01-108294).
HYDR IM C51wd Operator's Manual
=600) is intended to reduce the number of viable microorganisms