4. Installation Instructions

To open the wash chamber door, pull the ring
inside the chemical door.
Tools and supplies required for installation:
• Slot screwdriver
• Channellocks
Ensure that HIP Cleaning Solution (instrument wash
chemical) is available. All other supplies are included with
the Hydrim unit.
Specifications:
Voltage:
220 – 240 V
Frequency:
50 Hz
Rated load:
2 kW
Circuit
10 A per
breaker:
phase
Utility Hook-ups
Hose / Cord
Length /
Diameter
Hot inlet
1.9 m
G 3/4"
Cold Inlet
1.9 m
G 3/4
Drain
1.5 m
3/4"
Electrical
1.8 m
AWG 18-3
* unit will function with water pressure down to 0.5 bar.
drain hose
cold water
inlet hose
hot water
inlet hose
HYDR IM C51wd Operator's Manual
Height:
475 mm
Width:
600 mm
Depth:
460 mm
Depth with
780 mm
door open:
Weight:
34 kg
Running Noise: 60 dBA
Max. Distance
Water
Distance from
Pressure
inlet / drain
(optimal)*
1.5 m
1-10 bar
1.5 m
1-10 bar
inlet valves
The electrical outlet
should not be located
directly behind the unit
and should be accessible.
The powercord must be
routed away from the
back panel and hotwater
inlet hose.
Installation Configurations:
If the Hydrim is installed in a Sterilisation Center,
the manufacturer of the Sterilisation Center should
allow a 75 mm space at the top, back and both
sides of the Hydrim. This will facilitate installation,
levelling, and service access to the Hydrim.
The cabinetry will need to be vented and allow the
air exhaust from the dryer to escape. At least 75mm
of space should be provided behind the rear panel.
Do not move the Hydrim into place by maneuvering
the open wash chamber door. This may cause to
door to become misaligned and leak.
Option 1. Cabinet / Sterilisation Center
Shut-off
valve
Yes
Option 2. Sliding Shelf
Yes
Option 3. Counter Top
Page 4