Schiller Argus VCM Mode D'emploi page 67

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

ARGUS VCM
SpO2 Alarm Limits
Alarm Limits determine the high and low points of patient data at which the monitor
will sound an alarm.
With the monitor in the normal monitoring mode:
1. Press the Alarm button 9 times until SpO2 high alarm limit is displayed.
2. Leave the limit unchanged or press the Up/Down (+/-) selection buttons as needed
to change the limit to another value.
3. Press the Alarm button once again until SpO2 low alarm limit is displayed.
4. Leave the limit unchanged or press the Up/Down (+/-) selection buttons as needed
to change the limit to another value.
Figure 38. SpO2 High Alarm Limit Setting
Figure 39. SpO2 Low Alarm Limit Setting
ARGUS VCM
Alarms and Limits
Operator's Manual
61

Publicité

Table des Matières
loading

Table des Matières