NIBP Monitoring
The AUTO indicator is flashing when s selected for the automatic mode interval. After the
first cycle, the cuff target pressure is not used; rather, the cuff inflates to a pressure level
20~30 mmHg above the previous systolic reading.
The automatic NIBP cycles continue until one of the following occurs:
Note: During Auto mode, if an NIBP limit violation alarm occurs, this alarm disables any
automatic NIBP measurement until the alarm is released.
Note: The interval is the time from the beginning of one measurement cycle to the beginning
of the next measurement cycle.
Note: The selected interval is displayed on the Auto cycle display. The countdown timer for
initiating next measurement is displayed on the Time display.
38
ARGUS VCM
The monitor reaches the 5-minute limit for a STAT measurement.
he monitor halts because the NIBP start/stop button is pressed.
T
The monitor halts because the Auto button is pressed.
The monitor halts because of an alarm, alert, or error condition.
The AUTO cycle is changed to '-'.
Figure 20. Auto Mode Setting (i.e.) – 15 minute Period
Figure 21. Auto Mode of Measurement
Operator's Manual