ARGUS VCM
Initiating STAT mode of NIBP operation
1. Press the Auto button. The latest selected interval is displayed.
2. Press the Up/Down (+/-) selection buttons to set STAT.
Upon selection, automatic measurement is activated and the initial measurement will
be made in 3 seconds after you select an interval.
With the selected interval STAT, the monitor takes repeated NIBP measurements for 5
minutes. Current cuff pressures are not dynamically displayed during a STAT reading. The
measurement displays the NIBP reading from the previous cycle until the current cycle
finishes. (Before the first cycle finishes, the display appears blank.)
Note: During STAT mode, if an alarm, alert, or error condition occurs, the STAT
measurement will be terminated.
Stopping Blood Pressure Measurements
1. Press the NIBP start/stop button at any time that you wish to stop the current
measurement and deflate the cuff. If AUTO or Stat mode is underway, the mode
including interval time will be reset.
Note: During AUTO mode, pressing the NIBP start/stop button at the time before the next
auto measurement starts will cancel the AUTO mode and will be made a single blood
pressure measurement (MANUAL mode).
Figure 22. STAT Mode Setting
ARGUS VCM
NIBP Monitoring
Operator's Manual
39