Temperature Monitoring
Rectal Measurement
1. Remove the temperature probe from the probe holder.
The temperature probe runs a self-test, displaying '888.8' for a few seconds. When it is
ready for use, the monitor clears the temperature display. "rEC" appears in the
temperature display.
2. Load a probe cover onto the probe.
3. Apply a thin coat of water-based lubricant to the probe tip.
4. Separate the patient's buttocks with one hand.
5. Insert the probe tip 1 centimeter (3/8-inch) inside the rectal sphincter. Tilt the probe
slightly to ensure good tissue contact, and keep the buttocks separated throughout the
duration of the measurement.
The monitor displays the temperature reading for one minute.
Note: Use only a red-capped probe for rectal temperatures.
Note: During the measurement period, the temperature window displays a "walking
box"—a box with the sides illuminated sequentially. When the measurement is
complete, the monitor sounds a tone and displays the measurement in the
temperature window.
6. Remove the probe.
7. Eject the probe cover by pressing the ejection button, and hygienically dispose of it.
8. Return the probe into the probe well.
If a probe position error occurs during the temperature determination, the temperature
display alternates between the measured temperature and "P".