Fitting The Finger - marposs E86N Manuel D'installation Et D'utilisation

Detecteur touch a transmission radio
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

ENGLISH

13. FITTING THE FINGER

The following is a description of the procedure for fitting the finger on the probe mounted on a standard transmitter, but since the
compact transmitter incorporates the probe and the finger holder is identical, the procedure is the same for both types of transmitters.
To fit the finger on the probe proceed as follows:
Note: The procedure described below involves use of the crash protection pin. Said pin, inserted between the finger and
the probe, is designed to protect the probe in the event of accidental collisions affecting the finger (the probe finger
is detached).
-
Insert the crash protection pin (A) on the probe (B).
-
Use the wrench (supplied) to hold the probe finger holder (C) still and tighten the crash protection pin (A) with the other wrench.
Insert the pin (A) tightening wrench in the lower part of the pin - position (D).
-
Screw the finger (F) onto the crash protection pin (A).
Use one wrench to hold the pin still and the other to tighten the finger.
-
-
Insert the retainer (E), sliding it along the finger until it surrounds the crash protection pin (A). The retainer is designed to hold the
finger if the pin breaks.
-
In applications with radio transmission we recommend the use of styluses
-
Which are ceramic rather than made of steel, to avoid any (unlikely) interference. If the crash protection pin is not used, screw the
finger directly onto the probe; using one wrench to hold the finger holder (C) still and the other to tighten the finger.
Figure 13-1
86
A
B
F
A
A
D
E
E86N- Touch system with radio transmission
C

Publicité

Table des Matières
loading

Table des Matières