Immunity test
38
Voltage dips, short
interruptions and
voltage variations on
power supply input
lines
IEC 61000-4-11
Power frequency
(50/60 Hz) magnetic
field
IEC 61000-4-8
NOTE: UT is the a.c. mains voltage prior to application of the test level.
IEC 60601 test level
0 % UT for 0.5 cycle
at 0°, 45°, 90°, 135°,
180°, 225°, 270°, and
315°
0 % UT for 1 cycle
at 0°
70 % UT
for 25 cycles
at 50 Hz at 0°
for 30 cycles
at 60 Hz at 0°
0 % UT for 250 cycles
at 50 Hz at 0°
for 300 cycles
at 60 Hz at 0°
30 A/m
Compliance level
0 % UT for 0.5 cycle
at 0°, 45°, 90°, 135°,
180°, 225°, 270°,
and 315°
0 % UT for 1 cycle
at 0°
70 % UT
for 25 cycles
at 50 Hz at 0°
for 30 cycles
at 60 Hz at 0°
0 % UTfor 250 cycles
at 50 Hz
at 0°for 300 cycles at
60 Hz at 0°
30 A/m
Electromagnetic
environment –
guidance
Mains power quality
should be that of a
typical commercial or
hospital environment.
If the user of the
breast pump requires
continued operation
during power mains
interruptions, it is
recommended that
the breast pump be
powered from an
uninterruptible power
supply or a battery.
It may be necessary
to position the breast
pump further from
sources of power
frequency magnetic
fields or to install
magnetic shielding.
The power frequency
magnetic field should
be measured in the
intended installation
location to assure that it
is sufficiently low.