Guidance and manufacturer's declaration – electromagnetic immunity
This breast pump is intended for use in the electromagnetic environment specified below.
The customer or the user of the breast pump should assure that it is used in such an environment.
The Magic InBra™ breast pump was tested for immunity to electromagnetic disturbances and
passed using the following criteria:
1. No change in vacuum level and operation of the breast pump.
2. The pump can be controlled.
3. The Bluetooth® communication may be impaired temporarily but may never be completely
interrupted in such a way that a new pairing is necessary.
Reset (The following sequence can be used to reset the device in the event of an EM Disturbance)
1. Ensure the breast pump is not attached to the breast.
2. Ensure the device is powered off.
3. Disconnect the power cable.
4. Disconnect the power adapter from the wall outlet.
5. Press and hold the Power and the Start/Pause button for 10 seconds.
6. Wait 10 seconds.
7. Press the Power button for 1 second until the lights turn on to power on the device.
Immunity test
Electrostatic discharge
(ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
IEC 60601
test level
+/- 8 kV
contact discharge
+/- 15 kV
air discharge
± 2 kV
for power supply lines
± 0,5 kV, ± 1 kV
Line-to-line
Compliance level
+/- 8 kV
contact discharge
+/- 15 kV
air discharge
± 2 kV
for power supply lines
± 0,5 kV, ± 1 kV
Line-to-line
Electromagnetic
environment –
guidance
Floors should be
wood, concrete or
ceramic tile. If floors
are covered with
synthetic material,
the relative humidity
should be at least 5%.
Mains power quality
should be that of a
typical commercial or
hospital environment.
Mains power quality
should be that of a
typical commercial or
hospital environment.
37