Table 4
For ME EQUIPMENT and ME SYSTEMS that are not LIFE-SUPPORTING
Recommended separation distances between portable and mobile RF communications equipment and the BC30
The BC30 is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the
BC30 can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equip-
ment (transmitters) and the BC30 as recommended below, according to the maximum output power of the communications equipment.
Rated maximum output
power of transmitter
W
0,01
0,1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be determined
using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W)
according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures,
objects and people.
Separation distance according to frequency of transmitter
150 kHz to 80 MHz
d = 3.5
V
1
0.117
0.370
1.170
3.700
11.7
114
(m)
80 MHz to 800 MHz
d = 3.5
E
1
0.1167
0.3689
1.1667
3.6893
11.6667
800 MHz to 2.5 GHz
d = 7
E
1
0.2334
0.7378
2.3334
7.3786
23.3334