Guidance and Manufacturer's Declaration – Electromagnetic Emissions
The Inogen One® G2 Oxygen Concentrator is intended for use in the electromagnetic
environment specified below. The customer or the user of the Inogen One® G2 Oxygen
Concentrator should assure that it is used in such an environment.
Emissions Test
Compliance
RF emissions
Group 1
CISPR 11
RF emissions
Class B
CISPR 11
Harmonic Emissions
Class A
IEC 61000-3-2
Voltage fluctuations /
Complies
flicker emissions
IEC 61000-3-3
40
Electromagnetic Environment - Guidance
The Inogen One® G2 Oxygen
Concentrator uses RF energy only for
its internal function. Therefore its RF
emissions are very low and not likely
to cause any interference in nearby
equipment.
The Inogen One® G2 Oxygen
Concentrator is suitable for use in all
establishments including domestic
establishments and those directly
connected to the public low-voltage
power supply network that supplies
buildings used for domestic purposes.