Low Priority Alerts
The following low priority alert messages are accompanied by a double beep and a
solid red light.
Message Display & Text
Condition/Action/Explanation
External
External power supply is too low to run the unit, which
Power Low
is running on battery power despite being plugged in.
Check external power supply connections. If condition
persists, contact your equipment provider.
External
External power supply is too high. Check external power
Power High
supply connections. If condition persists, contact your
equipment provider.
Battery Low
Battery power is low, with less than 10% remaining. Attach
Attach Plug
external power supply or power down and insert a fully
charged battery.
Battery Error
Battery error has occurred. Switch to a new battery or
See Manual
remove battery and operate concentrator using external
power supply. If battery error recurs with same battery,
stop using the battery and contact your equipment
provider.
Oxygen Low
Concentrator is producing oxygen at a slightly low level
See Manual
(<82%). If condition persists, contact your equipment
provider.
24
Low Priority Alerts (continued)
Condition/Action/Explanation
Message Display & Text
Battery has exceeded its charging temperature and charging
Remove Battery
has stopped. The battery will not charge while this alert
to Cool
is present but will begin to charge when the battery
temperature returns to the normal operating range.
If battery charging is desired sooner, remove the battery
from the concentrator and allow it to cool in an open area for
approximately 10-15 minutes. Then, re-insert the battery into
the Inogen One® G2. If the problem still persists, contact your
equipment provider.
Concentrator is producing oxygen but cannot report
Comm Error
See Manual
battery status. Replace battery. If condition persists,
contact your equipment provider.
Service Needed
The concentrator requires servicing at the earliest
convenience. The concentrator is operating to specification
and may continue to be used. Contact your equipment
provider to arrange for service.
The concentrator's oxygen sensor has malfunctioned.
02 Sensor Fail
See Manual
You may continue to use the concentrator. If the condition
persists, contact your equipment provider.
25
Audible and Visible Signals Chapter 4