Selects dissolved or gaseous measurement phase. Refer to
on page 19.
Start automatic data acquisition. Refer to
Start memory storage view .Refer to
Starts continuous purge mode. Refer to
Display program identification information
Once you have completed the installation and start-up procedures defined in
the analyzer can be operated independently, making measurements as a portable carbon dioxide
gas analyzer.
You may store these measurement values for later analysis via the Hach 3654 program (on your
personal computer) or the memory view mode (on the instrument). The Hach 3654 program
operation is described in detail throughout the manual, where relevant.

5.2 Taking measurements

To start the analyzer, press the POWER switch (located bottom left of the keyboard). When you turn
power on, the instrument displays the software version number briefly, and then starts a series of
start-up tests. This process should take only a few minutes, during which time a [tSt] message is
displayed on the LCD. Once the tests are completed, a clearing message [---] is displayed for a
further sixty seconds before the instrument switches automatically to measurement mode.
Note: The start-up tests will be repeated until successful, or for a maximum of 15 minutes. If the tests are still
unsuccessful after 15 minutes, a [Err] message will be displayed and the instrument locked. Turn the instrument off
and retry the start-up procedures, but if the problem persists, please contact your local Hach Lange service
representative for assistance.
Measurement updates are displayed every 20 seconds, and it is normal for the right-most digits to
vary in reaction to slight variations in gas content. For accurate measurements, the sensor's
membrane must be at the same temperature as the sample to be analyzed. If this not the case, allow
some sample to pass through the flow chamber for about 3 minutes before taking any
measurements.
The LCD includes a right-side marker to distinguish between gas concentration measurements and
temperature. This marker also indicates the measurement display units (k/kg, V/V, etc. depending on
the instrument configuration).
To switch between gas measurement and temperature measurement, press the Up/Down Arrow
buttons. To illuminate the LCD for approximately three minutes, press the Backlight button.
5.2.1 Select gas measurement phase
Ensure the gas measurement phase is correct for this measurement. For details on viewing the
current setting, refer to
setting, refer to
Gas measurement phase
5.2.2 Measurement mode
You can set the measurement mode to normal or maximum. For details on how to do this refer to
Measurement mode
on page 21.
14 English
Reviewing instrument configuration
on page 19.
Automatic data acquisition
Viewing stored measurements
Continuous purge
on page 19. For details on changing this
Gas measurement phase
on page 15.
on page 16.
on page 29.
Installation
on page 8,