2.4 Pipetting
Identification
Your pipette is supplied with autoclave-resistant stickers in four 
diffe  r ent colours. These can be attached to the surface above the 
finger rest to personalise/help identify your pipette (Fig. 2).

2.5 Safety Notes

Use the pipette only with tip(s) attached.
Make sure that no liquid enters the pipette.
Do not lay the pipette down when a pipette tip is fixed and
contains liquid.
Do not set the volume outside the specified range.
Do not use any aggressive solvents or acetone to clean the pipette.
Do not use cleaning agents in combination with autoclaving or
UV-Treatment.
Only use original parts and accessories.
Before pipetting aggressive chemicals or organic solvents, check
their compatibility with the pipette and pipette tips.
Avoid temperature differences between the pipette and pipette tip
as this could lead to incorrect volume dispensing.
Check the dispensing volume of liquids that strongly differ from
the physical properties of water, eg. with different density (see
also section 6 Calibration).
Each ErgoOne® pipette has its 
own serial number. This serial 
number can be found behind the 
tip ejector button 
pressed down.
Fig. 2
ErgoOne
®
 when it is 
2
User Manual
9