6. Calibration

During manufacture, ErgoOne® pipettes are aligned for transferring 
aqueous solutions and calibrated according to EN ISO 8655. We there-
fore also recommend that you regularly check your pipettes at defined 
intervals of, for example, 3 or 12 months. 
Generally, realignment is not necessary with normal use. However, 
if your pipettes still do not achieve the specified limiting values 
(see section 5 Technical Data), check the following points: 
Is the pipette seal broken? 
Is it dripping? (see section 3 Troubleshooting)
Is the density of the liquid you are using different from that of water?
Are there any temperature differences between the pipette, liquid
and air?
Was the liquid pipetted slowly and evenly?
If the check is unsuccessful, calibrate and realign your pipette. This 
can be carried out according to the following instructions. Alternatively, 
STARLAB offers a professional Pipette Calibration Service with a
fast turnaround.
Note
that each alignment means a zero point shifting. The amount
by which the pipette setting is changed is constant across the entire
measurement range. If for example 1000µl is realigned by 10µl
(1%) for a 100 –1000µl pipette, then for 100µl, the pipette must
also be corrected by 10µl = 10%.
Calibration & alignment
Calibration is a gravimetric analysis for which you require an analytical 
balance adjusted to the pipetting volume, a weighing vessel and 
distilled water (room temperature). 

General information

Only use TipOne
Same temperature for pipette, tip and water
Each tip must be pre-wetted with distilled water at least three times.
Always pipette evenly and slowly and hold the pipette upright.
Immerse no more than 3 mm of 
the pipette tip into the liquid when taking a sample.
Position the pipette at an angle of 30 – 45° to the inner wall of the 
tube when dispensing liquid.
For multi-channel pipettes, all channels must be checked separately.
 pipette tips and attach these securely.
®