2.2 Setting the Volume

Open the volume adjustment catch by 
pulling up the volume control button 
(Fig. 1), until you hear a click. Turn the 
volume control button 
reduce the aspirated volume, or counter 
clockwise to increase it. The set volume is 
shown in the four-step volume display 
from top to bottom, with a white line indi-
cating the decimal point. When increasing 
the volume setting, pass the required 
value by 1/3 of a turn and then slowly de-
crease to reach the volume, making sure 
not to go beyond the required volume. 
After you have finished setting the volume, 
push the volume control button 
its lock position until you hear a click. The 
pipette is now ready to use. 
Volume range ErgoOne®
ErgoOne® model
Volume range
2.5µl
10µl
20µl (micro cone)
20µl (standard cone)
100µl
200µl
300µl
1000µl
5000µl
Examples of typical volume displays:
10µl Pipette
10.00µl
6
ErgoOne
 clockwise to 
1
 back to 
1
Colour code
(min./ max.)
0.1 – 2.5µl
Red
0.5 – 10µl
Red
2 – 20µl
Red
2 – 20µl
Yellow
10 – 100µl
Yellow
20 – 200µl
Yellow
30 – 300µl
Green
100 – 1000µl
Blue
500 – 5000µl
Violet
200µl Pipette
0.50µl
200.0µl
User Manual
®
1
2
4
4
Fig. 1
Please note the minimum 
and maximum volume 
range of your pipette 
(Table 1). You have 
reached the end point of 
the volume adjustment 
when it becomes more 
difficult to turn the button 
and you can hear 
a clicking sound. 
Do not use excessive
force!
1000µl Pipette
20.0µl
1000µl
1
100µl