Troubleshooting

Impedance Issue with the Stimulator

Issue
Out of range ("Bad") impedance
at a follow-up or programming
visit
Presentation
Red impedance icon(s) shows on the
Patient Device screen when the CP is
connected to the Stimulator
Resolution
Press the impedance button to refresh impedance values.
If the bad impedance is still present, check that the
Stimulator is connected to the trial cable (Basic Trial
Cable or Percutaneous Extension). If using a ground pad,
confirm the ground pad is connected to the patient and
to the trial cable. Refresh the impedance values.
If the bad impedance is still present, reprogram the
stimulation using electrodes with in-range impedance
values.
If impedance is still bad on all electrodes, device
replacement may be necessary. Contact Axonics.
65