Programming the Stimulator Stimulation Settings
1
2
3
5
4
Adjusting Parameters
Several adjustable stimulation parameters are shown in the
"Stimulation Parameters" box ( 5 ).
To adjust frequency and pulse width:
Press the left and right arrows to the sides of the current value
to incrementally decrease and increase the parameter.
Frequency: the stimulation frequency can be adjusted from
2-130 Hz. Adjustments are made in 1 Hz increments from 2-50
Hz and 5 Hz increments from 50-130 Hz.
Pulse width can be set between 60 and 450 μs, adjustable in
10 μs increments