Viewing the Status of a Stimulator
Stimulator Errors
1
If a Stimulator error is present, a "Query Fault" button will
appear to the right of the "Disconnect" button in the lower right
corner of the Patient Device screen.
2
When the "Query Fault" button is pressed, a pop-up will appear
( 1 ) with a description of the error ( 2 ).
To resume the communication session and try to resolve the
error with reprogramming, press "Clear Error" ( 3 ) to remove
3
4
the error notification from the Stimulator. If additional errors
exist, the pop-up will
display information about the next error, which will also need to
be cleared.
Press "Close" to exit the pop-up ( 4 ). If the pop-up is exited
without clearing the errors, the ability to program the Stimulator
may be impaired.