Programming the Stimulator Stimulation Settings
52
To adjust ramping:
Ramping settings are active when settings have been set to
the Stimulator.
Note: The ramping settings are not active during test stimulation. To test
ramping, save the stimulation settings to the Stimulator by pressing "Set
Device Therapy".
Press the button for the parameter and a pop-up will appear
for adjusting the ramping setting.
Allows stimulation to ramp up and down between zero
amplitude and the targeted stimulation amplitude when
stimulation turns on and off.
Ramping time can be programmed at "Off", 1, 2, 4, 8, 15,
or 30 seconds.