Programming the Stimulator Stimulation Settings
Electrode Configuration
1
The CP provides the ability to use automatically generated
recommended electrode configurations or to manually set which
2
electrodes are active and inactive during stimulation.
Electrode Recommendations
The Programming screen displays up to four recommended
electrode configurations ( 2 ). These recommendations are
intended to provide directional guidance regarding which
electrode configurations to test. For tined lead stimulation,
the recommendations are generated based on the stimulation
thresholds and motor and sensory responses recorded
during tined lead placement or subsequent redefining of the
thresholds. Ultimately, electrode configuration selection should
be based on patient comfort and symptom reduction. To set the
electrode configuration to a recommended configuration, press
the image of the recommendation.
Note: PNE lead electrode recommendations reflect all available electrode
configurations.
Note: If tined lead thresholds are not saved to the Stimulator, no electrode
recommendations are provided.