Viewing the Status of a Stimulator

The Patient Device screen for a Stimulator that has already been
set-up and programmed will include the following information
1
about the status of the device:
2
Device information ( 1 ): This section includes basic information
about the Stimulator, including patient ID, Remote Control
ID, and Trial Start Date. It also contains an estimate of how
long Stimulator battery can deliver stimulation at the current
stimulation settings. Next to "Battery Life" the CP displays
the expected battery life when stimulation is delivered at the
programmed (or "base") amplitude and when stimulation
is delivered at the maximum amplitude (or "max"). This
information should be provided to the patient when their device
is programmed.
Lead impedances ( 2 ): An icon showing the impedance status of
each electrode is shown to the right of each electrode. See manual
section Understanding Impedance Values for more information
about the qualitative impedance indicators.
The CP automatically checks impedance when it connects to the
Stimulator. To check the impedances again, press the impedance
button:
.
For more information on lead impedances, see Checking the
lead connection in manual section Setting up a New Stimulator.