Define Tined Lead Thresholds
Manually Recording the Motor and Sensory Responses
6
Motor and sensory responses associated with
stimulation at each electrode can be manually recorded.
Entering the motor and sensory responses is optional.
2
4
For example, if the patient is not awake during the
implant procedure, the sensory response can be left as
"None."
1
If saved, the response data and stimulation thresholds
5
8
are displayed when communicating with a Stimulator.
This data can be used to inform which stimulation
7
settings to use for therapy, and the CP will provide
3
recommended electrode configurations based on
this information. For example, if the stimulation with
a particular electrode elicited a "bad" sensation,
that electrode will not be included in recommended
electrode configurations.
Clear All Thresholds
The 'Clear All' button -
( 7 ) - appears once a
threshold value has been obtained (at the first test
stimulation).
Pressing this button will clear established thresholds
on all electrodes and allows this process to be started
over. This option may be useful if the tined lead is
repositioned, for example.