Configuration - Dräger X-am 5600 Notice D'utilisation

Masquer les pouces Voir aussi pour X-am 5600:
Table des Matières

Publicité

Marking offset channels:
Feature
ToxicTwins (X-am 5000 / 5600)
CO H
compensation (X-am 5000 / 5600)
2
H
offsetting (X-am 5600)
2
For more information on the features, see the technical manual:

Configuration

NOTICE
Only trained and competent users are permitted to change the instrument
configuration.
To individually configure an instrument with standard configuration, the instrument
must be connected to a PC via the USB infrared cable (order no. 83 17 409) or the
E-Cal system. The configuration work is carried out using the PC software
Dräger CC Vision. The PC software Dräger CC Vision can be downloaded from
the following web address free of charge: www.draeger.com/software.
Change configuration: see Technical Manual.
The following infrared sensors can be plugged into the gas detection instrument:
DrägerSensor Dual IR Ex/CO
2
DrägerSensor IR Ex (order no. 6812180)
DrägerSensor IR CO
(order no. 6812190)
2
DrägerSensor Dual IR Ex/CO
2
DrägerSensor IR Ex ES (order no. 6851881)
DrägerSensor IR CO
ES (order no. 6851882)
2
To simplify matters, this document generally refers to gases (e.g. Ex/CO
IR Ex). This refers to all existing sensor types.
Standard instrument configuration:
1)
Dräger X-am 5600
2)
Bump test mode
2 3)
Fresh air calibration
4)
Hydrogen offset
2) 5)
Operating signal
Capture range
2)
Switch off
16
Reading on the display
HCN+
CO+
ch
+
4
(order no. 6811960)
ES (order no. 6851880)
Extended bump test
On
On
On
On
allowed
2)
LEL factor
ch
4
H
2
2) 6) 7)
STEL
(short-term average)
2) 7) 8)
TWA
(shift average)
9)
Alarm A1
9)
Alarm A1 at O
sensor
2
9)
Alarm A2
®
1)
X-am
is a registered trademark of Dräger.
2)
Different settings can be selected to meet customer requirements on delivery. The current
setting can be checked and changed with the Dräger CC Vision PC software.
3)
The fresh-gas adjustment/zero-point adjustment is not supported by the CO
infrared sensor and the XXS O
4)
In the case of activated XXS H
5)
A periodic short flashing indicates the operating capacity of the instrument. If there is no
operating signal, correct operation cannot be guaranteed.
6)
STEL: average value of an exposure over a short period, generally 15 minutes.
7)
Interpretation only if the sensor is designed for this.
8)
TWA: shift averages are workplace limit values for generally eight hours per day of exposure for
five days a week during a working life.
9)
Latching and acknowledgement of alarms A1 and A2 can be configured with the Dräger CC
Vision PC software.
Activation of the H
offset adds the LEL gas concentration of the activated XXS
2
H
HC to the LEL gas concentration of the activated DUAL IR Ex/CO
2
activated IR Ex and it is output to the display instead of the IR Ex display.
or
2
Indicated by "+" in the display.
Previously set alarm thresholds are retained, so the IR Ex-channel alarm is
triggered earlier in some case in the presence of hydrogen (H
Selecting or disabling the capture ranges (only applies for the measuring
mode):
The capture range is selected in the measuring mode (factory setting) and
permanently disabled in calibration mode.
The CC-Vision PC software can be used to select or disable the capture ranges
for the measuring mode.
4.4 vol. % (4.4 vol. % corresponds to 100 %LEL)
4.0 vol. % (4.0 vol. % corresponds to 100 %LEL)
STEL function - disabled
Average value duration = 15 minutes
TWA function - disabled
Average value duration = 8 hours
can be acknowledged, non-latching, pre-alarm,
rising flank
cannot be acknowledged, latching, like main
alarm, falling flank
cannot be acknowledged, latching, main alarm,
rising flank
2
.
3
HC and activated Ex channel in the DUAL IR Ex/CO
2
NOTICE
).
2
channel of the
or IR Ex.
2
or the
2

Publicité

Table des Matières
loading

Table des Matières