If the calibration is not successful:
The mobile app will display the calibration as failed.
Repeat the calibration and have the sensor unit tested by maintenance
personnel as appropriate.
Check the test gas and follow the instructions in the mobile app.
For more information, see the technical manual.
5.4

Charging the sensor unit

WARNING
Risk of explosion!
The chargers are not constructed in accordance with regulations for wet
weather conditions and explosion protection.
► Do not charge in explosion-hazard areas!
► Use the supplied battery charger (type GTM96180-1217.9-5.9, order no.
6850018) with magnetic connector to charge the sensor unit.
► Ambient temperature during the charging process: 41 °F to 95 °F (+5 to
+35 °C).
WARNING
Health hazard!
Magnets can affect the function of pacemakers and implantable cardioverter
defibrillators (e.g. actuation of the reed switch).
► Observe a minimum distance of 7,87" (20 cm) between the magnet
connection and the implanted devices to avoid malfunctions and health
hazards.
The rechargeable battery cannot be changed. If the rechargeable battery is
faulty, it can be replaced during maintenance.
The sensor unit can be charged whether turned on or off.
When turned on and connected, the charging status of the sensor unit can
be displayed in the mobile app.
When turned off, the status LED pulses during the charging process (
When the sensor unit is fully charged, the status LED also flashes twice
between pulses (
).
Dräger X-pid
).
®
8x00 Sensor Unit
The typical charging time is 5 hours from complete discharge to full charge.
Never store the device for extended periods (max. 2 months) without a power
supply to avoid a complete discharge of the battery.
5.5

Configuration

Only trained personnel are permitted to carry out modifications to the device
configuration.
The sensor unit can be configured in conjunction with the Dräger X-pid xx00
mobile app, see technical manual.
5.6

Replacing the water and dust filter

Replace the water and dust filter if contaminated or blocked.
Only use the water and dust filters listed as spare parts.
Designation and description
Water and dust filter
(without Luer adapter)
Water and dust filter set
(incl. Luer adapter for OD ~5 mm and
3 mm)
5.7

Cleaning

The sensor unit does not require any special care.
● In the case of heavy contamination, the sensor unit can be cleaned with cold
water. A sponge can be used for wiping if necessary.