Instrument Settings; Preparations For Operation; Sensor Settings - Dräger X-am 5600 Notice D'utilisation

Masquer les pouces Voir aussi pour X-am 5600:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Instrument settings

The following parameters can be changed on an instrument:
Name
Password
1)
Operating signal LED
1)
Horn operating signal
Switch-off mode
2)
Shift length (TWA)
Short-term exposure level (STEL)
1)
At least one of the two operating signals must be switched on.
2)
Corresponds to averaging time and is used to calculate the TWA exposure value.
3)
Interpretation only if the sensor is designed for this.
4)
Corresponds to averaging time and is used to calculate the STEL exposure value.

Sensor settings

The following parameters can be changed on the sensors:
Name
Alarm threshold A1 (in measuring unit)
Alarm threshold A2 (in measuring unit)
1)
Interpretation type
Alarm threshold STEL
1)
(in measuring unit)
Alarm threshold TWA
1)
(in measuring unit)
1)
Interpretation only if the sensor is designed for this.
Check of parameters
To ensure that the values were correctly transferred to the gas detection
instrument:
Press the data from X-am 1/2/5x00 in Dräger CC Vision.
Check parameters.
Range
Numerical range (3-digit)
Yes / No
Yes / No
"switch-off allowed" or
"switch-off prohibited" or
"switch-off prohibited at A2"
60 - 14400 (in minutes)
(Setting for exposure alarm)
3) 4)
0 - 15 (in minutes)
(Setting for exposure alarm)
Range
0 - A2
A1 – measuring range end value
Inactive, TWA, STEL, TWA+STEL
0 – measuring range end value
0 – measuring range end value
Operation

Preparations for operation

Before using the instrument for the first time, insert a charged NiMH T4
power pack or batteries approved by Dräger see "Replacing the batteries /
rechargeable batteries" on Page 9.
The instrument is now ready for operation.
To reduce the risk of ignition of a flammable or explosive atmosphere, strictly
adhere to the following warning statements:
Only use power pack types ABT 01xx, HBT 00xx or HBT 01xx. See the marking
on the rechargeable battery for permitted rechargeable batteries and the
corresponding temperature class.
Substitution of components may impair intrinsic safety.
Remove rubber boot only if the Dräger X-am 5600 is used in combination with
the accessory or if used in the Dräger X-zone 5000/5500 (AAC 00xx).
WARNING
00133322.eps
5