Battery pre-alarm

Intermittent alarm:
Flashing special symbol »
« on the right side of the display:
Acknowledge pre-alarm:
Press the [ OK ] key, only the audible alarm and the vibration alarm are
switched off.
The battery still lasts approx. 20 minutes after the first battery pre-alarm.

Battery main alarm

Intermittent alarm:
Flashing special symbol »
« on the right side of the display:
The battery main alarm cannot be acknowledged or cancelled:
The instrument is automatically switched off again after 10 seconds.
Before the instrument is switched off, the visual, audible and vibration
alarms are activated for a short time.

Instrument alarm

Intermittent alarm:
Special symbol »
« displayed on the right side of the display:
The instrument is not ready for operation.
Commission maintenance personnel or the Dräger Service Centre to
eliminate the error.
Activating the Info mode
In measuring mode, press the [ OK ] key for approx. 3 seconds.
If any warning or fault messages exist, the corresponding note or error codes
are displayed (see Technical Handbook). Press the [ OK ] key successively
for the next display. The peak values and the exposure values TWA and
STEV will be displayed.
If no key is pressed for 10 seconds, the instrument returns automatically to
measuring mode.