14. Electromagnetic compatibility

Guidance and manufacturer's declaration– Electromagnetic emissions
L'
R16M is suitable for use in the specified electromagnetic enviroment. The purchaser or user of the R16M
should assure that is used in an electromagnetic environment as described below:
Emissions test
RF emissions
CISPR11
RF emissions
CISPR11
Hermonic emissions
IEC/EN 61000-3-2
Voltage fluctuations/flicker emissions
IEC 61000-3-3
Guidance and manufacturer's declaration– Electromagnetic immunity
This R16M is suitable for use in the specified electromagnetic environment. The purchaser or user of the
R16M should assure that it is used in an electromagnetic environment as described below:
Immunity test
Electrostatic
discharge (ESD)
IEC/EN 61000-4-2
Radiated
electromagnetic
field
IEC 61000-4-3
Electrical fast
transient/burst
IEC/EN 61000-4-4
Surge IEC 61000-4-5
Conducted
disturbances
induced by RF fields
IEC 61000-4-6
Compliance Electromagnetic environment
Group 1
Class B
Complies
Complies
IEC 60601-1-2
Test level
8 kV contact
2/4/8/15 kV air
3 V7m
80 MHz to 2.7 GHz
2kV for power supply
lines
1 kV for input/output
lines >3m
0.5/1 kV differential
mode
0.5/1/2 kV common
mode
3V
150 kHz to 80 MHz
R16M User's Manual – Page 141
This R16M uses RF energy only for its internal function.
Therefore, the RF emission is very low and not likely to
cause any interference in nearby electronic equipment.
This R16M is suitable for use in all establishments,
including domestic establishments and those directly
connected to the public low-voltage power supply
network that supplies buildings used for domestic
purposes.
Compliance
Electromagnetic Environment
level
Floors should be wood, concrete or
IEC
60601-1-2
ceramic tile. If floors are covered with
Test level
synthetic material, the relative humidity
should be at least 30%
Portable and mobile FR communications
IEC
60601-1-2
equipment should be used no closer to
Test level
any part of the R15C including cables.
Minimum distance 30 cm.
Mains power quality should be that of a
IEC
60601-1-2
typical
Test level
environment.
Mains power quality should be that of a
IEC
60601-1-2
typical
Test level
environment.
Portable and mobile RF communications
IEC
60601-1-2
equipment should be used no closer to
Test level
any part of the R15C including cables.
Minimum distance 30 cm.
commercial
or
hospital
commercial
or
hospital